Today's dose rates: Leivonmäki 0.122 µSv/h More
You are here: Home > Services > Measurement Services > Calibration, testing and irradiation services > Ionising radiation > Calibration, testing and irradiation services in Radiation Metrology Laboratory

Calibration, testing and irradiation services in Radiation Metrology Laboratory

OperationQuantityDosemeterRadiation type and average energyMeasurement rangeMeasurement uncertainty (*)
Calibration

Reference air kerma rate



Brachytherapy well ionization chambers



Gamma radiation:
192Ir (375 keV) and
125I (28 keV)

0,1 µGy/h - 
50 mGy/h

2 - 6 %

Calibration

Dose-area product

Dose-area product meters in
X-ray diagnostics

X radiation:
50 - 150 kV

10 mGy cm2 -
10 Gy cm2

5 %

CalibrationDose-length productDose-length product meters in X-ray diagnostics

X radiation:
100 - 150 kV

1 mGy cm - 500 mGy cm

5 %

Calibration

X-ray tube voltage

X-ray tube voltage measurement instruments

X radiation:
10 - 150 kV

10 - 150 kV

n. 0,5 kV

Calibration or irradiation

Absorbed dose or dose rate in tissue

Personal dosemeters,
radiation protection measurement instruments

Beta radiation:
90Sr/90Y (0,80 MeV)
85Kr (251 keV)
147Pm  (62 keV)

0,06 mGy/h -
1,3 Gy/h

4 - 6 %

Check of dose ratemeter (1)Ambient dose equivalent rateRadiation protection measurement instrumentsGamma radiation
60Co (1,250 MeV)
and 137Cs (662 keV)
20 µSv/h -
25 Sv/h
         5 %
Check of dose ratemeterAmbient dose equivalent rateRadiation protection measurement instrumentsNeutron radiation:
241Am - 9Be fast neutrons
30 µSv/h -
500 µSv/h
       4 - 6 %
Check of dose ratemeter(2)

Surface activity
Surface emission rate

Contamination measuring instrument 
(4)

Beta radiation:
14C (49 keV),
36Cl (251 keV), 90Sr/90Y (0,80 MeV)

Alfa radiation:
241Am (5,54 MeV)  (3)

ca 80 kBq/m2 -
110 kBq/m2 

ca 10 % 

(1) Testing of radiation measuring instruments for rescue operations and civil defence with 137Cs gamma radiation at the dose rate range 30 µSv/h - 30 mSv/h (Guide STUK VAL 1.3)
(2) Checking of external sensors for rescue operations and civil defence with 90Sr/90Y beta standard source (Guide STUK VAL 1.3)
(3) Maximum energy
(*) Extended uncertainty with a coverage factor of 2.

Page updated 06/06/2009